H. Hilgenkamp, J. Mannhart, et al.
IEEE TAS
We have developed a small scanning tunneling microscope (STM) to be incorporated into a scanning electron microscope (SEM). Vibration isolation and damping is achieved solely with Viton dampers. As a stand-alone unit, a tunnel-gap stability of about 1 Å is reached at atmospheric air pressure without additional sound protection. Stability improves by at least an order of magnitude when incorporated into a SEM.
H. Hilgenkamp, J. Mannhart, et al.
IEEE TAS
D. Anselmetti, Ch. Gerber, et al.
Review of Scientific Instruments
G. Binnig, H. Rohrer
Surface Science
M. Bammerlin, R. Lüthi, et al.
Applied Physics A: Materials Science and Processing